Japan HIOKI IR4057-50 insulation high resistance meter original company goods (FKH4057-50)
Japan HIOKI IR4057-50 insulation high resistance meter original company goods (FKH4057-50)
HIOKI insulation high resistance meter/high resistance meter adopts an integrated design.
IR4057-50 is an insulation high-resistance meter with 5 ranges and a test voltage of 50V/100MΩ~1000V/4000MΩ. It has powerful functions and fast response speed, which helps shorten on-site working time.
- Supports wireless communication: transmits measurement results to smartphones or tablet devices, and allows you to combine test results with on-site photos or drawings to quickly create reports (when used with wireless adapter Z3210), thus helping to shorten the time on site Insulation measurement working time
- Measurement data will be directly input into the Excel form through wireless transmission (this function is expected to be included in the 2021 version upgrade)
- 5 test ranges from 50 V / 100MΩ to 1000 V / 4000MΩ
- digital scale bar
- Stable and high-speed digital display, PASS/FAIL judgment response time is only 0.3 seconds
- Can withstand the impact of a 1m drop on concrete
- High-brightness LED display screen, test line with LED light switch (option L9788-11 or L9788-10)
- 200mA test continuous check, can measure AC/DC voltage
- Suitable for testing solar power systems and electric vehicles
IR4057-50 insulation resistance meter, supports wireless communication
When used with the wireless adapter Z3210 (Z3210 is an optional accessory and needs to be purchased separately), the measurement results can be transmitted to a smartphone or tablet device, and allows you to combine the test results with on-site photos or drawings to quickly create reports .
For a detailed introduction to Z3210, please click here .
Product specifications:
Test voltage | 50VDC | 125VDC | 250V DC | 500VDC | 1000V DC |
Effective maximum display value | 100 MΩ | 250 MΩ | 500 MΩ | 2000 MΩ | 4000 MΩ |
Accuracy 1st effective measurement range MΩ |
±4% rdg. 0.200-10.00 |
±4% rdg. 0.200-25.0 |
±4% rdg. 0.200-50.0 |
±4% rdg. 0.200-500 |
±4% rdg. 0.200-1000 |
Minimum measured resistance value | 0.05 MΩ | 0.125 MΩ | 0.25 MΩ | 0.5 MΩ | 1 MΩ |
Overvoltage protection |
AC 600V (10s) |
AC 660V (10s) | |||
DC voltage |
4.2V (resolution 0.001V)~600V (resolution 1V), 4 ranges,
Accuracy: ±1.3% rdg. ±4 dgt., Input resistance: 100 kΩ or more
|
||||
AC voltage |
420V (resolution 0.1V) / 600V (resolution 1V), 2 ranges, 50/60 Hz,
Accuracy: ±2.3% rdg. ±8 dgt., Input resistance: 100 kΩ or more, average value
|
||||
low resistance | Used to check the continuity of the ground wire, 10 Ω (resolution 0.01 Ω)~1000 Ω (resolution 1 Ω), 3 ranges, basic accuracy: ±3 % rdg. ±2 dgt., test current 200 mA or more ( When 6Ω or less) | ||||
show | FSTN LCD with backlight, digital scale bar | ||||
response time | PASS/FAIL judgment takes about 0.3s (under the conditions specified by HIOKI) | ||||
Other functions | Display MΩ measurement value after 1 minute, live circuit alarm, automatic discharge, automatic DC/AC check, comparator, anti-fall, automatic power saving | ||||
power supply |
No. 3 (LR6) alkaline batteries x 4, continuous use: 20 hours (based on internal testing)
Number of measurements: 1000 times (5sON, 25sOFF loop, insulation measurement from the lowest measured resistance value to maintaining rated measurement voltage)
|
||||
Volume and weight | 159 mm W × 177 mm H × 53 mmD, 640 g (including battery and test leads) | ||||
appendix | Connecting cable L4930 × 1, alligator clip L4935 × 1, test pin L4938 × 1, lanyard × 1, instruction manual × 1, size 3 (LR6) alkaline batteries × 4 |